Skip to content

Commit

Permalink
iio: temperature: mlx90632 Relax the compatibility check
Browse files Browse the repository at this point in the history
Register EE_VERSION contains mixture of calibration information and DSP
version. So far, because calibrations were definite, the driver
compatibility depended on whole contents, but in the newer production
process the calibration part changes. Because of that, value in EE_VERSION
will be changed and to avoid that calibration value is same as DSP version
the MSB in calibration part was fixed to 1.
That means existing calibrations (medical and consumer) will now have
hex values (bits 8 to 15) of 83 and 84 respectively. Driver compatibility
should be based only on DSP version part of the EE_VERSION (bits 0 to 7)
register.

Signed-off-by: Crt Mori <cmo@melexis.com>
Cc: <Stable@vger.kernel.org>
Signed-off-by: Jonathan Cameron <Jonathan.Cameron@huawei.com>
  • Loading branch information
Crt Mori authored and Jonathan Cameron committed May 27, 2019
1 parent bce0d57 commit 389fc70
Showing 1 changed file with 7 additions and 2 deletions.
9 changes: 7 additions & 2 deletions drivers/iio/temperature/mlx90632.c
Original file line number Diff line number Diff line change
Expand Up @@ -81,6 +81,8 @@
/* Magic constants */
#define MLX90632_ID_MEDICAL 0x0105 /* EEPROM DSPv5 Medical device id */
#define MLX90632_ID_CONSUMER 0x0205 /* EEPROM DSPv5 Consumer device id */
#define MLX90632_DSP_VERSION 5 /* DSP version */
#define MLX90632_DSP_MASK GENMASK(7, 0) /* DSP version in EE_VERSION */
#define MLX90632_RESET_CMD 0x0006 /* Reset sensor (address or global) */
#define MLX90632_REF_12 12LL /**< ResCtrlRef value of Ch 1 or Ch 2 */
#define MLX90632_REF_3 12LL /**< ResCtrlRef value of Channel 3 */
Expand Down Expand Up @@ -667,10 +669,13 @@ static int mlx90632_probe(struct i2c_client *client,
} else if (read == MLX90632_ID_CONSUMER) {
dev_dbg(&client->dev,
"Detected Consumer EEPROM calibration %x\n", read);
} else if ((read & MLX90632_DSP_MASK) == MLX90632_DSP_VERSION) {
dev_dbg(&client->dev,
"Detected Unknown EEPROM calibration %x\n", read);
} else {
dev_err(&client->dev,
"EEPROM version mismatch %x (expected %x or %x)\n",
read, MLX90632_ID_CONSUMER, MLX90632_ID_MEDICAL);
"Wrong DSP version %x (expected %x)\n",
read, MLX90632_DSP_VERSION);
return -EPROTONOSUPPORT;
}

Expand Down